This web page is designed to be a basic introduction to flow cytometry. It summarises a lecture given by Dr Hazel Davey to third year undergraduates at Aberystwyth University. The lecture forms part of the Microbial Physiology, Biochemistry & Biotechnology [BS33920] module. The material presented covers the following topics:
Cytometry is a technique for making measurements (-metry) on cells (cyto-). Cytometric measurements are usually made via microscopic methods.
FLOW cytometry refers to the same kind of measurements, however the measurements are performed whilst the cells flow past an array of optical detectors.
Flow cytometry involves the same types of measurements that could be made by (fluorescence) microscopy:
Light of one
wavelength is absorbed and light of a different wavelength is emitted. Emitted
light is always of lower energy and therefore longer wavelength than the
exciting light. In the diagram on the left, blue light interacts with a cell.
Much of the light is scattered by the cell surface or cytoplasmic inclusions.
However, some of the light is absorbed by a fluorescent compound within the
cell and re-emitted as red light. The change in wavelength allows fluorescence
to be distinguished from light scatter
Only a brief, non-technical description of how flow cytometers work will be given. There are several components that make up a flow cytometer these include:
Only the fluidics and optics will be discussed in more detail here.
As mentioned earlier, the important
characteristic of flow cytometry is that measurements are made as the
cell sample flows through the instrument. If you simply used a tube to deliver
the sample, the cells (represented by black dots in the diagram on the left)
will not be delivered reproducibly to the measuring point (yellow dot). If a
narrower tube was used to ensure reproducible sample delivery it would be
likely that blockages would result. To overcome these problems, a method
known as hydrodynamic focusing is used. This involves
introducing a slowly moving stream of cells into a quickly moving carrier
fluid. The carrier (or sheath) fluid constrains the cells to the centre of the
tube and thus the cells are reproducibly delivered to the measuring point.
The optical system of a flow cytometer is composed of one or more light sources, together with a series of lenses, filters and detectors. The lenses and filters act to deliver the light source to the measuring point and also to transfer scattered light and fluorescence from the measuring point to the detectors.
A variety of different light sources are currently used in commercial flow cytometers. These include:
Optical filters allow light of selected wavelengths to pass through while limiting transmission of other wavelengths. The two most commonly used in flow cytometry are :
Dichroic filters are selective mirrors which allow transmission of long wavelengths while reflecting short wavelengths. In the example shown in the diagram on the left, cells were illuminated with 488 nm light from an Argon ion laser. This has resulted in scattering at the incident wavelength and fluorescence at longer wavelengths. By using a 500 nm dichroic the majority of the scattered light is reflected downwards to one detector, whilst the majority of the fluorescent light is transmitted to another detector.
Band pass filters allow light of a specific wavelength, or narrow band of wavelengths, to pass through. In the example shown in the diagram on the left, a mixed light source resulting from scattering and fluorescence reaches a 630 nm band pass filter. This filter is designed to allow transmission of a narrow band of wavelengths between 620 and 640 nm. In this way fluorescence from a single cellular component or added stain can be isolated and quantified.
The optics and fluidics combine to provide a method of delivering the sample in a reproducible manner to the focused light source. A series of filters are then used to split the resulting light into bands representing scattering by the cells and fluorescence from different compounds within them.
In order to appreciate the value of flow cytometry it is first of all necessary to consider what alternative techniques are available. As mentioned earlier, flow cytometry essentially allows one to make the same sorts of measurements that could be made by conventional light or fluorescence microscopy. However, conventional microscopy has the following disadvantages:
Alternatively, in some cases, measurements may be made using bulk biochemical techniques. However, rather than measuring individual cells measurements reflect average values for whole populations of cells.
In comparison to these techniques flow cytometry offers several advantages:
Moldovan, A. (1934): Photo-electric technique for the counting of microscopical cells. Science 80, 188-189.
In 1934 Moldovan described the first instrument that could be described as a flow cytometer, although the term flow cytometer was not coined until much later. The instrument consisted of a glass capillary tube mounted on microscope stage. Initially Moldovan used narrow tubes but found that the cells tended to block them, whilst the edges of the tubes sometimes came into the field of view and caused interference with the measurements. When wider tubes were used cells were not delivered reproducibly and errors in detection and measurement were observed.
Gucker, F.T. et al. (1947): A photoelectric counter for colloidal particles. Am. J. Chem. 69, 2422-2431.
The photoelectric counter was designed during WW2 primarily to test the efficiency of gas mask filters against particles. In the majority of experiments dust particles were used but Bacillus spores were also analysed. The device used filtered air to carry and constrain the sample rather than the sheath fluid used in modern instruments. While the air had the same effect of reproducibly delivering the sample, it obviously limited the types of biological samples that could be analysed in this device. The light source used in Gucker's instrument was the headlight from a Ford car, yet this was sufficient to detect spores. In the conclusions to Gucker's paper it was stated that "The principle (of flow cytometry) should have wide application in (...) bacteriology", however over 30 years were to pass before serious work began in microbial flow cytometry.
Crosland-Taylor, P.J. (1953): A device for counting small particles
suspended in a fluid through a tube. Nature 171, 37-38.
Coulter
(1953): US Patent
A device developed by Crossland-Taylor in 1953 introduced the concept of hydrodynamic focusing for reproducible delivery of cells suspended in a fluid. Using this device accurate counts of blood cells were obtained. In the same year Wallace Coulter patented the first of several devices for counting blood cells. The Coulter counter immediately superseded the Crossland-Taylor device and is still widely used today. The Coulter principle involves the detection of individual cells via electrical rather than optical means as they flow through a small orifice.
Kamentsky, L.A., et al. (1965): Spectrophotometer: new instrument for ultrarapid cell analysis. Science 150, 630-631.
In 1965 Kamentsky and colleagues described an instrument with a custom quartz flow cell which could be used for making measurements on a biological parameter. Previous systems had used flow devices only to detect and count cells. Using this device absorption of light at 254 nm was used to estimate the nucleic acid content of indiviual cells at rates of up to 500 cells/sec. Light scattering was also measured simultaneously to indicate the size of the cells with good reproducibility between samples.
Fulwyler, M.J. (1965): Electronic separation of biological cells by volume. Science 150, 910-911.
Also in 1965, the Coulter principle was used in the first demonstration of cell sorting technology. The technology used to build the cell sorter came from an advance in a very different field. The deflection of electrostatically charged ink drops had recently been used in ink-jet printing and this technique was adapted to sort droplets containing cells. The Coulter volume was used to distinguish between mouse and human red blood cells (mean volumes 50 and 100 µm respetively) and then these were separated from each other with high purity. Separation of hybridoma cells from cell culture debris was also demonstrated. It was shown that the sorting process did not adversely affect the cells (viability >96% in all cases) and that the separated cells grew with an identical generation time to an untreated control.
The first commercial flow cytometer reached the marketplace in 1969. It was called the ICP 11 and was distributed by Phywe, Göttingen. A picture of this and other early flow cytometers can be seen at the Partec flow museum. This may be compared with a more modern flow cytometric cell sorting instrument shown in the photograph on the right. The instrument shown is a Coulter Epics Elite and was acquired by the Institute of Biological Sciences, University of Wales, Aberystwyth in August 1993. The instrument is equipped with 4 lasers and can collect data for 8 different cellular parameters. In addition it can sort selected fractions or individuals into tubes, microtitre plates or onto the surface of agar plates.
Since the 1960's flow cytometric applications involving the study of mammalian cells have steadily increased. A variety of cell types have been analysed. Since flow cytometry requires cells to be suspended individually in fluid, blood cells are amenable to analysis with minimal pretreatment. There is growing interest in using flow cytometry routinely for the analysis of clinical specimens. Antibodies specific for different CD markers may be labelled with fluorescent tags allowing white cell typing to be carried out. This approach may be used to identify and type leukaemias or to monitor the progress of diseases such as HIV/AIDS. Other applications include analysis of cells isolated from tumors. DNA stains are used to detect aneuploidy which is indicative of cancerous cell growth. In the research environment, flow cytometry is frequently used to study apoptosis (programmed cell death). Multiparametric flow cytometry enables researchers to monitor the biochemical changes that occur within cells as they enter an apoptotic state.
Despite the early predictions of Gucker et al. and the obvious benefits that a technique like flow cytometry has for microbial research, the first microbial flow cytometry papers did not apppear until the late 1970's. Between 1977 and 1980 there were a few papers published which generally demonstrated that flow cytometry could be used to detect microorganisms of various types.
During the 1980's Steen and coworkers demonstrated the utility of flow cytometry for studying the cell cycle of bacteria. Prior to this work, knowledge of the bacterial cell cycle was based on the work of Cooper and Helmsetter who studied division in synchronised cells in the late 1960's. By using flow cytometry to look at individual cells, Steen and coworkers were able to use unsynchronised and thus unperturbed populations. During the last 15 years various new microbiological applications of flow cytometry have been developed and this has led to an increase in the number of microbial flow cytometry papers. The graph shown illustrates the increase in use of flow cytometry in microbiology. The graph was produced from the results of a year by year Web of Knowledge search of article titles, keywords and abstracts with the following query strings:
The value returned for the first search was used for the total flow cytometry papers published in a given year, the value for the second search was used as the microbial flow cytometry paper total. It should be noted that records prior to 1991 did not contain abstracts.
A variety of applications of flow cytometry in microbiological research have been developed and published. These include:
Viability can be monitored via a variety of methods:
Non fluorescent substrates such as FDA are also used to indicate enzyme activity and can act as gene reporters in a similar manner to colorimetric assays involving lacZ. By using fluorescent reporters enzyme activity in individual cells can be monitored leading to a growing role for flow cytometry in microbial molecular biology. Monitoring the heterogeneity of gene expression within a population is much more powerful than studying population-based responses - particularly when one considers that cells may respond in different ways to environmental stimulii according to their position in the cell cycle.
It was mentioned above that fluorescein liberated from probes such as FDA tend to leak out of cells. A new compound was recently described (Zlokarnik et al. (1998): Quantitation of transcription and clonal selection of single living cells with β-lactamase as reporter. Science 279, 84-88.) which overcomes this problem.
The compound described, CCF2/AM, is a non-fluorescent precursor. It is sufficiently non-polar to easily cross the cell membrane where it is converted to fluorescent CCF2 by non-specific esterase action. CCF2 is polar and is thus trapped within the cell. β -lactamase then cleaves CCF2 into two separate fluorescent molecules with different emission and excitation wavelengths that can be independently monitored by flow cytometry.
An important fluorescent reporter molecule for flow cytometry is the so-called green fluorescent protein (GFP). This was originally isolated from the jellyfish Aequorea victoria (see photo). Production of this protein is now used as a gene reporter in both mammalian and bacterial systems. The original green protein has been modified by mutations to produce proteins which fluoresce at different wavelengths, giving a range of spectrally discrete reporters including RedFP, CyanFP, BlueFP, SapphireFP and YellowFP. This development has enabled multiparametric research into the expression of several different genes. Thus gene activity can be monitored for different genes at the same time in 1000s of individual cells.
The take home message from this article may be summarised in the following two sentences:
The information provided on this and other pages by me is my own personal responsibility and not that of the University of Wales, Aberystwyth. Similarly, any opinions expressed are my own and are in no way to be taken as those of U.W.A.
There have been
accesses to this page since 30th November 2000.